1.4. Line Failure Detection
z
IC2020 is able to automatically detect if
the telephone line is installed properly.
z
Line failure detection starts after IC2020 is
powered on for one minuite, and repeats
every 15 sec thereafter.
z
If line failure is detected, IC2020 will
respond as 4 steps illustructed below:
1. If line faliure lasts over 10 minuites,
IC2020 will emit a beep sound and the Red
LED will flash 3 times every 3 seconds.
2. If line failure lasts under 10 minutes and
either Emergency or Nurse button is
pressed, IC2020 will emit a vocie prompt
“
Phone line troulble
” and the Red LED will
flash 3 times every 3 secounds.
3. The voice promot “Phone line trouble” will
be repeated as defined by Command 53
(please refer to section
3.2. Programming
Your IC2020
)
4. If you hold the Nurse button for 20
seconds, a long beep will be heared and
then release the Nurse Button. The
repearting cycle of voice prompt “
Phone
line trouble
” will be reset into 6 hours.
z
During the line failure period, if the
Emergency or Nurse button is pressed or
any sensor is triggered, the red LED will
flash 3 times every 3 seconds and the
voice prompt “
Phone line trouble
” will be
played. The reporting will then be made
after the line failure is restored.
z
If line failure is detected and the device
low battery reporting has been previously
made before the line failure period, then
no more device low battery reporting will
be made after the line failure is restored.
However, if the device low battery
condition is first detected during the line
failure period, then a device low battery
reporting will be made after the line failure
is restored.
z
During the reporting period after line
failure restore (if the Emergency or Nurse
button is pressed or any sensor is
triggered during the line failure period),
line failure detection will be temporarily
stopped.
z
Once line failure is restored for one
minute or more, the red LED will dim and
the voice prompt will also be stopped. If
line failure happens again, it will repeat
step 1 to step 4 as illustructed in the item
3 of
1.4 Line Failure Detection
.
z
Line failure and restore will not be
reported.
z
At least one telephone and account #
must exist in order for line failure
detection to start. Line failure detection
will not start if no telephone nor account #
exist (even after powered on for one
minute).
1.5. How to install the Control
Panel
Locating a suitable position for the
Control Panel
z
The Control Panel requires main power
and a constant telephone connection.
z
The Control Panel should be easily
accessible.
z
The Control Panel should not be placed in
a damp location, such as a bathroom.
z
The Control Panel should not be placed
close to any heat source, such as
microwave ovens, which can reduce
signal strength.
z
The Control Panel should not be located
alongside other radio transmitting devices,
such as mobile phones, cordless phone,
or wireless computer network (Wi-Fi)
devices.
z
Important: When drilling into a wall, ensure
there are no hidden cables or pipes.
Mounting the Control Panel
The Control Panel can be mounted on the wall
or wherever desired (e.g. on the table). Ensure
the Control Panel is fitted at approximately
chest height where the buttons, microphone,
and speaker can be easily accessed and
operated.
z
Use the 2 holes of the Wall Mounting
Bracket as a template, mark off the holes’
positions.
z
Drill 2 holes and insert the wall plugs if
fixing into plaster or brick.
z
Screw the base to the wall.
z
Hook the Control unit onto the Wall
Mounting Bracket (holding the unit with
the front facing you).