16
)
When Mobility Timer expires and no
reset signal is received, IC2020 starts
to emit one long beep every 5 minutes
to alert the user. After 30 minutes, if
reset signal is not received, a mobility
alarm reporting will be made.
)
The reporting method for mobility
alarm is the same as the emergency
alarm. The report will be sent based on
your reporting methods without Guard
Time waiting.
)
When (
0
) is selected, no code is sent
to Monitoring Centers.
)
When IC2020 is in idle mode, press
and hold the White NURSE button for
16 seconds, until you hear a long
beep. Then release the button; the
Mobility function will be disabled.
Command 45
Follow-On Mode Choice
z
This Command is only available for Digital
reporting with Command #38
set as (
0
) or
(
2
).
z
It is used to select the Follow-On Choice
for IC2020 after it has made a report to the
Monitoring Center.
z
After reporting, IC2020 can wait for further
commands, or enter a Two-Way Full
Duplex Communication, or Listen-In Half
Duplex Communication immediately
according to the setting under this
command:
Function
Selection
Result
(0)
WAIT
COMMAND
(1) LISTEN
(2) TWO
WAY
z
Factory Default is
set
as (
2
),
TWO WAY
.
<
<
N
N
O
O
T
T
E
E
>
>
)
When (
0
)
is selected, after reporting the
Contact ID Code, IC2020 will wait for 1-
60-min, set by Cammand 46, and then
follows the DTMF commands that are
received from the Central Monitoirng
Station for action.
If no command is received after the
waiting period, the system will
automatically return to Idle mode..
)
When (
1
)
is selected, after IC2020 has
made a report to the Monitoring Center,
it will automatically enter
Listen-In Half
Duplex
for 1-60 minutes (set by
Command 46) allowing the Monitoring
Center Personnel to listen in to what is
happening on the other side.
)
When (
2
) is selected, after reporting,
the IC2020 will immediately enter a
Two-way Full-duplex Communication
mode for 1-60 minutes (set by
Command 46), that gives the user the
ability to speak directly with the
Monitoring Center Personnel.
)
If either Scancom or Tunstall reporting
format is selected, IC2020 will ignore
Command #45 selections. It will always
open a Two-way voice for Follow-On
choice.
Command 46
Follow-On Mode & Two way Period
z
This command is used to program the
allowable time duration for Follow-On
mode choice (Commend #45) & Two-way
voice communication period (Speech
mode). It can be set from (
01
) (1 minute) -
(
60
) (60 minutes).
Function Selection
Result
(01) 1
minute
(02) 2
minutes
(03)-(60)
3-60 minutes
z
Factory default is set as (
10
),
10 minutes
.
The allowable time duration for either
Wait
Command
,
Two-way
,
Listen-In Only
or
Two-way voice comminucation
in
Speech mode
and
Nurse call
is 60 mins.
During the two-way voice communication
period, IC2020 will emit two beeps via the
telephone handset to alert the call recipient
the remaining time of 20 & 10 sec.
Pressing the assigned DTMF key can
extend another 1-60 minute(s).
Содержание IC2020
Страница 1: ...l 02 April 2010 For INSTANT CARE...