Cell Connections – Fuses in the Cell Cable
51
If the one or both fuses are open, all the current readings will be near zero, and a red
CA Overload
indication may be seen at the bottom of the runner window.
If this test indicates an open fuse, use the procedures described above to check
both
fuses. This test cannot tell
which fuse is blown. Both fuses can blow at the same time.
If the fuse test indicates an open fuse, and the fuses both check out good with an ohmmeter, some other
problem has occurred in the cables or the instrument. Contact Technical Support at Gamry Instruments at your
earliest convenience.