Calibration – Cable Calibration
41
4)
Open Gamry Framework™ software. Select
Experiment
>
Named Script...
The
Select a script to run
window appears. From the list of scripts, choose
calcable.exp
, then
click the
Open
button.
5)
The
Cable Capacitance Calibration
window appears.
In the
Cable Tag
field, enter a unique name for the cable you are calibrating.
Choose the desired
Action
radio button:
To calibrate the cable, choose
Cal Cable
.
To reset the vales to zero (if, say, the calibration doesn’t work), choose
Zero Values
.
Click the
OK
button.
6)
The
Performance Tips
window appears.
Make sure that all of the tips are true, then click the
OK
button.