![Gamry Instruments Reference 3000 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/gamry-instruments/reference-3000/reference-3000_operators-manual_3393579067.webp)
Auxiliary Electrometer Option – Connections Using Custom Cables
67
5
Analog Input
Even Ch +
Red
Purple
High side of even channel
working side
6
Ground
Floating Ground --
--
7
Digital input
!Cable
Ground if cable present
8
Digital In
!Power On
Ground if AE powered
9
Ground
Floating Ground
Note that pins 8 and 9
must
be grounded on any cable that is used to take voltage readings.
The analog inputs are both labeled (with labels on the wires) and color coded. The wires are black for odd-
numbered channels (Channel 1, 3, 5, and 7) and red for the even-numbered channels (Channels 2, 4, 6, and
8).
The working side plug (the positive input) is purple, and the reference side (the negative input) is yellow.
Connections Using Custom Cables
You can build custom cables for AE connections. Follow the pin designations in the table above. Pins 7 and 8
must be grounded.
AE Specifications
See Appendix E: Auxiliary Electrometer Specifications