Command
code
Description
Parameter Response Data/response interpretation
0x79
Set TSP
configuration
T, M, N,
X.XE-XX,
Y.YE-YY,
D, P
T is TSP
M is cycle period in minutes (time
between fires)
N is number of cycles to run
X.XE-XX is high pressure limit
Y.YE-YY is low pressure limit
D is on time in seconds
(sublimation duration)
P is pressure window
(0 for on, 1 for off/ignore)
If only T parameter is specified,
the current TSP configuration is
returned
0x82
Get TSP pressure
high limit
T
X.XE-XX
where T is TSP
X.XE-XX is current high pressure
limit
0x8B
Set TSP pressure
supply
T, S
where T is TSP and S is supply
0x8C
Get TSP pressure
supply
T
S
where T is TSP and S is supply
0x8F
Schedule
firmware
upgrade
Send command and cycle power.
Upon power up, the loader mode
will run
0xDE
Get TSP status
T, 0
YY
where T is TSP
YY is two digit status code:
00 for TSP Initializing
01 for TSP Off
02 for TSP Armed (Running)
03 for TSP Armed (Waiting,
pressure window, supply is off)
04 for TSP Armed (Waiting,
pressure window, pump pressure
outside the defined pressure
window)
05 for TSP Armed (Waiting,
Interlock not satisfied)
06 for TSP Armed (Waiting, Other
TSP is firing)
07 for TSP Firing
0xDF
Get TSP active
filament
T
X
where T is TSP
X is the active filament
Page 47
900031_J
900031_J - Serial communication