Command
code
Description
Parameter Response Data/response interpretation
0x28
Turn off TSP
T
where T is TSP
0x29
Set TSP active
filament
T, X
where T is TSP and X is filament
number (1-4)
0x2B
Clear TSP
filament status
T
where T is TSP
Clears status for all TSP filaments
0x2D
Turn on TSP
T
where T is TSP
0x30
Get TSP
sublimation
target
T
X, P
where T is TSP
X is value in Amps/Watts and P is
‘A’ for Amps or ‘W’ for Watts
0x31
Get TSP pressure
low limit
T
X.XE-XX
where T is TSP
X.XE-XX is current pressure low
limit
0x33
Set supply auto
restart
S, X
where S is supply and X is ‘Y’ for
on and ‘N’ for off
0x34
Get supply auto
restart
S
‘YES’ or
‘NO’
where S is supply
0x37
Turn on HV
S
where S is supply
0x38
Turn off HV
S
where S is supply
0x3B
Get/set setpoint
Setpoints (N
parameter) 1-4
are relays, 5-8
are digital
outputs
N,F,S,
X.XE-XX,
Y.YE-YY
N,F,S,
X.XE-XX,
Y.YE-YY,
A
Where:
N is the set-point number (1-8)
F is the function (0 for Off, 1 for
Pressure Setpoint, 2 for HV Error,
and 3 for HV On/Off Indicator)
S is supply (1-4)
X.XE-XX is On Pressure, Y.YE-YY is
Off Pressure*
A indicates: 1 = Relay/TTL output
energized, 0 = Relay/TTL output
de-energized
*Off Pressure must be at least
20% greater than On Pressure. If
only N parameter is specified, the
current setpoint configuration
and state is returned.
Page 44
900031_J
900031_J - Serial communication