Command
code
Description
Parameter Response Data/response interpretation
0xE0
Get TSP filament
status
T
T,X,X,X,X
where T is TSP and X is filament
status (0 for good, 1 for open/
questionable)
0xE0
Get TSP filament
status
T
T,X,X,X,X
where T is TSP and X is filament
status (0 for good, 1 for open/
questionable)
0xE1
Get TSP filament
auto advance
mode
T
T, X
where T is TSP
X is current filament auto advance
mode (0 for off, 1 for next, and 2
for rotate)
0xEA
Set TSP pressure
high limit
T, X.XE-XX
where T is TSP and X.XE-XX is new
high pressure limit
0xEB
Set TSP pressure
low limit
T, X.XE-XX
where T is TSP and X.XE-XX is new
pressure low limit
0xED
Get/set pump
name
S, TEXT
TEXT
where S is supply and TEXT is
pump name (maximum 15
characters)
TEXT is the current pump name.
If only S parameter is specified,
the current pump name is
returned
0xFF
System restart
X
where X is the restart mode (0 for
restart, 1 for restart and start
loader mode)
Page 48