The Output Configuration page is divided into two sections:
•
Analog Outputs,
•
Relay Outputs.
Analog Outputs
The Analog Outputs section of the Output Configuration page allows
for configuration of each of the three isolated analog outputs of the
ViscoSite transmitter. The Analog Outputs section of the Output
Configuration page is shown in Figure 5.18.
Figure 5.18: Analog Outputs Section (Output Configuration)
The parameters that can be edited are as follows:
•
Output Type
–
Allows selection of how the signal is transmitted on
an isolated analog output and the minimum and maximum output
levels for each configuration. The available choices for an analog
output are:
•
Current 4-20mA;
•
Voltage 2-10VDC;
•
Current 0-20mA;
•
Voltage 0-10VDC.
•
Measurement
–
Allows selection of which measurement parameter
is output on the selected isolated analog output channel. See Table
5.4 for the complete list of parameters that may be selected for
output on Analog outputs.
•
Min
–
Allows the user to configure the lower limit of the selected
measured parameter (0/4mA or 0/2VDC signal), depending on the
type of output selected. If the measured parameter drops below this
entered value, an alarm will be triggered.
•
Max
–
Allows the user to configure the upper limit of the selected
measured parameter (20mA or 10VDC signal), depending on the
type of output selected. If the parameter rises above this entered
value, an alarm will be triggered.
The values in the “Units” column are derived from the measurement
units selected in the
Display Setup
page (Section 5.6.2). They cannot be
edited from this screen.