3.11.2.1 Communications with the ViscoSite
Transmitter via USB using a Virtual COM
Port (Local PC)
•
Create a virtual COM port on your local Windows PC by
installing the (provided) ST Link virtual COM port
drivers or run another virtual COM port software
installer package
on all PC’s that will run the ViscoSite
software. Be sure to use the installer appropriate for
your Windows operating system version and select the
32-bit or 64-bit installer version that matches your
operating system.
•
Install the ViscoSite Software (provided) onto your
local
Windows PC. Follow all directions to complete the
installation.
•
Connect the
Male USB-A to Male USB-A
cable
(provided) between the ViscoSite transmitter and an
available USB port on the PC.
•
Power cycle the transmitter to ensure the serial output
of the transmitter are prepared properly for
communications. This will only be necessary once.