Securely fasten the cable to the transducer receptacle using the
threaded coupling at the end of the transducer cable. Figure 3.8
shows a representative transducer cable installation.
Figure 3.8: Connecting the Transducer Cable at the Transducer
Figure 3.9 indicates the pinout of the cable connecting receptacle on
the dome lid. Note that the pin layout is asymmetrical with Pin L at the
top. There is no Pin I.
Figure 3.9: Transducer Cable Connecting Receptacle Pin Layout (Dome
Lid, from Top)
If the provided cable is too long,
DO NOT attempt to cut it.
Cutting the
cable WILL affect the calibration of the system
and WILL yield
inaccurate measurements. Neatly wrap up the excess cable length.