The UNGUATOR® MB should not be pressed too deep in the UNGUATOR® Jar, thus the
lifting arm can reach the bayonet. For this reason we recommend that after screwing the
UNGUATOR® mixing unit on, the UNGUATOR® MB shaft should by manually drawn back to
the jar lid.
Error:
The UNGUATOR® jar cannot be screwed in the lifting arm because the shaft of the
UNGUATOR® MB seems to be too long.
Troubleshooting
: The UNGUATOR® MB must be replaced with the appropriate UNGUATOR®
MB with the correct shaft length. While the UNGUATOR® SMBs for 100 and 200 ml and for
the 300 and 500 ml jars have the same mixing blade diameter, their shaft length differs.
Error
: The lifting arm is unintentionally moved with the hand too far from the starting
position and the shaft is no longer aligned to the opening.
Troubleshooting
: The shaft can be moved by hand to the start position. The shaft touches
the hole and is lead into the opening.
Error
: The UNGUATOR® MB does not penetrate completely in the mixture, since the formula
is to pasty, firm, or in powder form.
Troubleshooting
: The reverse power of the lifting arm is from 3 to 5kp intentionally low to
avoid material damages. For this reason, the UNGUATOR® Q registers the resistance and
the return of the stroke direction. To ensure the complete mixing of the recipe, you can
for example, warm cold recipe substances in room temperature or assist pasty mixtures by
hand. Further help is usually no longer necessary.
UNGUATOR® Jar
Error
: The jar connected to the UNGUATOR® Q is not an UNGUATOR ® jar.
Troubleshooting
: The jar has to be substituted for an UNGUATOR® jar.
Error
: The UNGUATOR® jar lid is wrongly screwed on the UNGUATOR ® jar and
the UNGUATOR ® MB tilted at automatic hub.
Troubleshooting:
The UNGUATOR® jar should be unscrewed and correctly screwed together.
Error:
The UNGUATOR® jar bottom was not pressed fully down to the stop position prior to
weighing out or filling.
Troubleshooting:
Particularly lite components are not going to be accommodate in the
same size UNGUATOR® jar although the filling volume is in the average 40% more than the
32