Procedure:
1.
Enter the Laser Tools Page. This can be done by going to the Machine name, Diagnostics and then selecting Galvo
Calibration or by inputting: ID address /setup and then Setup.
2.
Lay out about 1 to 2 sheets of letter sized paper centered under the red dot. Ensure these sheets are flat
and smooth to produce the best calibration.
3.
Make sure your Galvo is focused if you have not already done so.
4.
Then go to the “Laser Tools Tab” and select “Run Lens Calibration” if you have not already done so.
*Refer to Lens Calibration on pg 39.*
5.
After, it passed go to Verify Camera Calibration and select the verify camera calibration option.
45
Section IX. Camera Functions
Focus Tag Calibration:
There may be times when the Focus Tag is damaged or is not read properly. It may be fixed by calibrating the
focus tag.
Make sure your machine is within focus before starting calibration. Write down the Z- stages coordinates to
ensure calibration is done correctly.
1.
Enter the Laser Tools page. Look for the Tag Autofocus Calibration section.
2.
Place the Focus tag in the center of the work space.
3.
Select the Take Snapshot button to perform a camera capture. The Focus Tag should appear in the
screen with a red box above the Tag. The Red box signifies that the Tag was detected. If the box does
not appear then the Tag may be damaged.