6. SpO
2
Monitor (LX-7230KM)
38
■
Applying Finger Clip Probe (SR-5C)
The SR-5C is suitable for middle/long-term measurement, and should be
clipped to the fingertip of an adult/ pediatric who is 30kg or above.
It should be positioned so that the LED (light source) is at the base of the
fingernail, as shown in the figure below. Use the finger mesh cover (FC-M) to
secure the probe. Also, use the finger mesh cover to get better measurement if
it is exposed to high-intensity light.
Light Source
Sensor
Proper attachment
Finger is not sufficiently
inserted.
Finger is inserted too far.
Measurement cannot be taken.
WARNING
Precautions for Use of Finger Clip Probe (SR-5C)
If continuously measuring over a long period of time, change the measuring
finger every 8 hours (recommended wearing hours) to prevent low-temperature
burn. Especially for continuous use on patient with peripheral circulatory
disturbance, change the measuring finger more frequently.
Do not use tape to fix the probe. It may cause edema or congestion.
CAUTION
Precautions for Use of Finger Clip Probe (SR-5C)
The probe should be positioned so that the light source is at the base of the
fingernail. Take care not to insert the finger too deep to prevent injury.
When removing the probe, do not pull on the cord.
Cover the probe with the finger mesh cover (FC-M) if it is exposed to high-
intensity light such as direct light and surgical light to prevent measurement error.
The SR-5C is designed for use on finger of an adult/ pediatric who is 30kg and
above. Do not use on other parts.
Before attaching the probe to a patient, clean the probe using a cloth moistened
with sterilizing alcohol.
Remove nail polish before taking measurements to enable stable
measurements.
It may not be possible to take measurements in case of an excessively high
hematocrit value. In such case, reattach the probe to a thinner finger.
Take note that it may not be possible to take measurements in place with
vibrations or while walking.
After attaching the probe, make sure that there is no failure message and the
pulse wave level on the LCD is upper than 2. If not upper than level 2, it may
measure incorrectly. In such case, warm the finger to improve blood circulation.
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...