13. Maintenance and Inspection
78
Daily Check List
No.
Inspected Date
Inspected by
Location
Device Type LX-7230N/KM
S/No.
Date of Purchase
Items Details
Criteria
Judgment
Appearance
Visually check for any damage,
cracks, chip, peeled label, and
loosen screw on the housing.
No abnormality should be
found.
□
OK/
□
NG
Visually check for the ring
condition of the battery
compartment lid.
No damage, kink, floating,
and adhesion of dust
should be found.
□
OK/
□
NG
Battery
Compartment
Visually check for the contact
springs, inside the LX-7230, to
the battery and the lock lever of
the battery compartment lid.
No deformation, cracks,
and rust should be found.
□
OK/
□
NG
Power Supply
Turn the power ON/OFF to verify
proper switch operation.
With batteries installed,
the LCD should turn ON.
□
OK/
□
NG
ECG Connectors
Visually check the connectors of
the cable and the LX-7230.
No damage, chip, and
adhesion of dust should
be found.
□
OK/
□
NG
ECG Lead cable
Visually check each lead for
damages.
No crack and damage
should be found.
□
OK/
□
NG
SpO
2
Relay Cable
(LX-7230KM only)
Visually check the cable, and
connector for damages.
No crack, chip and
damage should be found.
□
OK/
□
NG
SpO
2
Sensor
(Probe)
Visually check the cable,
optical
receiver, LED, and connector for
damages.
No crack, chip, damage,
and adhesion of dust
should be found.
□
OK/
□
NG
Wireless Channel
Verify whether the transmitting
channel and group ID are the
same with the receiving monitor.
Must match the wireless
channel check list.
□
OK/
□
NG
Transmission
Function
Turn the power ON and make
sure the information is displayed
on the receiving monitor.
Waveforms and values
should be received
without any problem.
□
OK/
□
NG
Display Function
Turn the power ON and verify
each display condition, such as
SpO
2
value and bargraph.
All data should be
properly displayed.
□
OK/
□
NG
Periodic Check
Check the date of the previous
periodic check.
Should be within one
year.
□
OK/
□
NG
Comment
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...