6. SpO
2
Monitoring (LX-7230N)
18
SpO
2
Monitoring (LX-7230N)
LX-7230N, which has a built-in Nellcor
®
SpO
2
module, is described in this
section.
Regarding the LX-7230KM, which has a built-in Konica Minolta
®
SpO
2
module,
refer to page 35.
< SpO
2
Sensors for LX-7230N >
The optional SpO
2
sensors available for LX-7230N are as follows.
The following table shows applicable patient and proper site for each SpO
2
sensor. Select the proper one depending on the purpose and intended use.
Sensor Types
Applicable Patient
Applied Site
OxiMax
®
MAX-I
Infant
(weight of 3 to 20Kg)
Toe
OxiMax
®
MAX-P
Pediatric
(weight of 10 to 50Kg)
Finger
OxiMax
®
MAX-A/AL
Adult
(weight of 30Kg and over)
Finger
Durasensor
®
DS-100A
Adult
(weight of 40Kg and over)
Finger
OxiMax
®
MAX-R
Adult
(weight of 50Kg and over)
Nose
OxiMax
®
MAX-FAST
Adult/Pediatric
(weight of 10Kg and over)
Forehead
Adult
(weight of 40Kg and over)
Finger
OxiMax
®
MAX-N
Neonate
(weight of less 3Kg)
Foot
WARNING
For SpO
2
monitoring, always use the sensor specified by Fukuda
Denshi. If any other sensor is used, high temperature rise of the sensor
may place the patient in danger of burns in the worst case.
As with all medical equipment, carefully route cables to reduce the
possibility of patient entanglement and strangulation.
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...