17
6. SpO
2
Monitoring
When the transmitter is used without the ECG lead cable, it will measure only
SpO
2
.
WARNING
When the SpO
2
probe (sensor) is in a connector-off condition, the SpO
2
alarm will not be generated on the receiving monitor. Make sure that the
SpO
2
probe (sensor)/ relay cable is securely connected. If the SpO
2
waveform/numeric data is not displayed, check the patient’s condition and
pay attention not to miss the connector-off condition.
CAUTION
When using the transmitter with only the SpO
2
sensor cable, ECG and
respiration measurements on the receiving monitor shall be turned off to
prevent an erroneous alarm.
There are varieties of SpO
2
sensors depending on the purpose and intended
use for LX-7230N and LX-7230KM. Select a proper one and attach it to the
patient.
CAUTION
SpO
2
sensor for LX-7230N and SpO
2
sensor for LX-7230KM are not
compatible. SpO
2
sensors for LX-7230KM cannot be used on the LX-
7230N transmitter; vice-versa. Make sure to use the correct sensor for
operation.
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...