32
4. SETUP AND MAINTENANCE
4
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0321 Focus Position Adjustment]
[Procedure]
Perform the focus position adjustment.
1. 135F/135H of the NC100AY: Install the mask to be adjusted
to the carrier.
2. Install the carrier.
3. MFC10AY: Install the mask to the carrier.
NOTE: Use a mount holder without film for the mount mask.
• Carrier ID, carrier type, mask type and position adjustment
value are displayed.
4. 135F/135H of the NC100AY: Select “135F” or “135H” for
“Mask type”.
135M/IX240M of the MFC10AY: Select “135M” or “240M” for
“Mask type”.
5. Except for mount mask: Insert a film into the carrier or mask.
NOTE: • For the NC100AY, use a film having more than five
frames.
• For the NC120AY, use a film having more than 200
mm in length.
6. Click the [Run] button.
• NC100AY: The “W-4827” message appears.
• MFC10AY: The “I-4877” message appears.
7. Confirm film insertion and click the [OK] button.
• The “I-4828” message appears.
• The position adjustment value is displayed when the
adjustment is completed.
8. Click the [Regist.] button.
• The focus position adjustment value is registered.
NOTE: The adjustment value is not registered if the [Regist.]
button is not clicked.
9. Repeat Steps 1 to 8 if there are other carriers or masks to be
registered.
10. Click the [Close] button.
• Operation returns to the menu screen.
SP3000_instruct_E.book 32 ページ 2007年12月11日 火曜日 午後1時6分