Chapter 6:
Replacement Procedures
69
Figure 6.28 Stand Head Replacement
Optional Equipment - Stand System
Sony UP-X898MD Printer
The Sony UP-X898MD thermal printer is a monochrome black and white printer which uses the USB 2.0
port to integrate with the Sonosite PX Stand. This is an optional device. The following instructions are for
any replacement needs if the Sony UP-X898MD has to be replaced.
Required Part
• P20006 Sony UP-X898MD B&W Printer
Required Tools
• #1 Phillips screwdriver
• 5mm hex bit with driver
Remove Printer
1
Remove the rear cover on the printer, exposing the USB and power cables per
Figure 6.29 on page 70
2
Disconnect the USB and power cables from the rear of the printer
3
Use a 5mm hex bit to loosen the two screws under the printer.
4
Lift the printer and mounting box from the stand.
5
Use a Phillips #1 bit to remove the 4 screws securing the printer to the mounting box.
6
Remove the Printer from the mounting box.
1
2
3
4
Содержание Sonosite PX System
Страница 1: ...i Sonosite PX Ultrasound System Service Manual ...
Страница 2: ...ii ...
Страница 4: ...iv ...
Страница 14: ...6 Chapter 1 Introduction ...
Страница 18: ...10 Chapter 2 Specifications ...
Страница 36: ...28 Chapter 3 Safety ...
Страница 46: ...38 Chapter 4 System Overview ...
Страница 54: ...46 Chapter 5 Troubleshooting ...
Страница 79: ...Chapter 6 Replacement Procedures 71 ...
Страница 80: ...72 Chapter 6 Replacement Procedures ...
Страница 82: ...74 Chapter 7 Maintenance ...
Страница 90: ...82 Chapter 8 Performance Testing ...
Страница 102: ...94 Appendix A Replacement Parts ...
Страница 109: ...101 ...
Страница 110: ...102 P25393 01 P25393 01 ...