![FujiFilm Sonosite PX System Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/fujifilm/sonosite-px-system/sonosite-px-system_service-manual_2342370022.webp)
14
Chapter 3:
Safety
Battery safety
To prevent the battery from bursting, igniting, or emitting fumes and causing personal injury or equipment
damage, observe the following precautions.
Caution:
Do not use solvents such as thinner or benzene, or abrasive cleaners on any
part of the system.
Caution:
Remove the battery from the system if the system is not likely to be used for a
month or more.
Caution:
Do not spill liquid on the system.
Caution:
Position the system to allow access to the mains power-cord connector.
WARNING:
The battery has a safety device. Do not disassemble or alter the battery.
WARNING:
Charge the batteries only when the ambient temperature is between 0° and
40°C (32° and 104°F).
WARNING:
Do not short-circuit the battery by directly connecting the positive and negative
terminals with metal objects.
WARNING:
Do not touch battery contacts.
WARNING:
Do not heat the battery or discard it in a fire.
WARNING:
Do not expose the battery to temperatures over 60°C (140°F). Keep it away
from fire and other heat sources.
WARNING:
Do not charge the battery near a heat source, such as a fire or heater.
WARNING:
Do not leave the battery in direct sunlight.
WARNING:
Do not pierce the battery with a sharp object, hit it, or step on it.
WARNING:
Do not use a damaged battery.
WARNING:
Do not solder a battery.
WARNING:
The polarity of the battery terminals is fixed and cannot be switched or
reversed. Ensure that the batteries are in the correct orientation.
WARNING:
Do not connect the battery to an electrical power outlet.
WARNING:
Do not continue recharging the battery if it does not recharge after two
successive six hour charging cycles. Replace the battery.
WARNING:
Do not ship a damaged battery without instructions from Fujifilm Sonosite
Technical Support. (See
“Introduction” on page 1
.)
WARNING:
If the battery leaks or emits an odor, remove it from all possible flammable
sources.
Caution:
To avoid the battery becoming damaged and causing equipment damage,
observe the following precautions:
• Do not immerse the battery in water or allow it to get wet.
• Do not put the battery into a microwave oven or pressurized container.
Содержание Sonosite PX System
Страница 1: ...i Sonosite PX Ultrasound System Service Manual ...
Страница 2: ...ii ...
Страница 4: ...iv ...
Страница 14: ...6 Chapter 1 Introduction ...
Страница 18: ...10 Chapter 2 Specifications ...
Страница 36: ...28 Chapter 3 Safety ...
Страница 46: ...38 Chapter 4 System Overview ...
Страница 54: ...46 Chapter 5 Troubleshooting ...
Страница 79: ...Chapter 6 Replacement Procedures 71 ...
Страница 80: ...72 Chapter 6 Replacement Procedures ...
Страница 82: ...74 Chapter 7 Maintenance ...
Страница 90: ...82 Chapter 8 Performance Testing ...
Страница 102: ...94 Appendix A Replacement Parts ...
Страница 109: ...101 ...
Страница 110: ...102 P25393 01 P25393 01 ...