58
Chapter 6:
Replacement Procedures
Figure 6.15 USB AUX Board
Power Button with ECG PCBA Replacement
Required Part
• P27479-XX Service Assembly, Power Button with ECG PCBA, Sonosite PX
Remove Power Button with ECG PCBA
1
Remove screw from the ECG Flex Interconnect Cable using a T8 bit per
Figure 6.16 on page 58
2
Disconnect the ECG Flex Interconnect Cable from the Main PCBA and Power Button with ECG PCBA
and remove.
3
Remove two screws from the Power Button with ECG PCBA and remove.
Figure 6.16 Power Button with ECG PCBA
Remove screws
Disconnect from Main
PCBA
Screw
(T8)
Screws
(T8)