50
Chapter 6:
Replacement Procedures
Figure 6.5 Front Panel Enclosure
5
Lift the User Interface from the system as shown in
Figure 6.6 on page 50
Caution:Do not lift the User Interface from the front. Two tabs on the back of the UI sit in recesses in the
Bottom Enclosure and this may break the tabs on the UI. Begin by lifting straight up until the tabs clear
the enclosure.
6
Once the tabs are clear, the rear of the UI may be placed on the Bottom Enclosure and the front lifted
up to access the interconnection cables.
7
Remove the UI to Main PCBA Cable and Persistent Control to Main Flex Cable from the Main PCBA.
8
Remove the UI from the system.
Figure 6.6 User Interface Removal
UI Pin
Upper Enclosure Recess
Remove UI to Main Cable
Remove Persistent Control to
Main Flex Cable