Chapter 3:
Safety
25
IEC 60601-2-37:2015 Medical Electrical Equipment - Part 2-37: Particular requirements for the basic safety
and essential performance of ultrasonic medical diagnostic and monitoring equipment (Edition 2.1)
IEC 60601-1-6:2013 Medical Electrical Equipment Part 1-6: General requirements for basic safety and
essential performance - Collateral Standard: Usability (Edition 3.1)
EMC standards classification
IEC 60601-1-2:2014, Medical Electrical Equipment, General Requirements for Basic Safety and Essential
Performance-Collateral Standard. Electromagnetic Compatibility. Requirements and Tests.
CISPR 11:2009, Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment Electromagnetic
Disturbance Characteristics—Limits and Methods of Measurement.
The Classification for the ultrasound system, docking system, accessories, and peripherals when
configured together: Group 1, Class A.
DICOM standard
Digital Imaging and Communications in Medicine (DICOM), Version 3.1, 2007 (NEMA).
The system conforms to the DICOM standard as specified in the Sonosite PX DICOM Conformance
Statement, available at
www.sonosite.com
. This statement provides information about the purpose,
characteristics, configuration, and specifications of the network connections supported by the system.
Security and privacy standards
The system includes security settings that help you to meet the applicable security requirements listed in
the HIPAA standard. Users are ultimately responsible for ensuring the security and protection of all
electronic protected health information collected, stored, reviewed, and transmitted on the system.
HIPAA:1996, 45 CFR Parts 160 and 164; Subparts A, C, and E, Health Insurance Portability
and Accountability Act (HIPAA) Privacy and Security Rules
NIST SP 800-53, Security and Privacy Controls for Federal Information Systems and
Organizations
Security and privacy
• The port for DICOM communication (specified by the user in the system settings; typically port 104,
2762,or 11112) is used for outgoing communication to the network.
• Anti-virus software is not installed on this device.
• This device has a single configurable listening port for the purposes of DICOM Echo and Storage
Commitment.
• The security and privacy-related configurable controls in Sonosite PX are:
• User roles and responsibilities
• Automatic user log off
• User authorization and authentication
• Data backup and recovery
• Data encryption (at rest and in transit)
• System and department administrators should follow the suggested technical and physical safeguards
listed below, as well as the detailed HIPAA guidelines to ensure HIPAA compliance:
•
Room Access Control:
Local procedures must be put in place to limit physical access to medical
equipment, to prevent accidental, casual, or deliberate contact by unauthorized individuals.
•
System Access Controls:
Access the system only through unique user accounts. Login
credentials must not be disclosed.
Содержание Sonosite PX System
Страница 1: ...i Sonosite PX Ultrasound System Service Manual ...
Страница 2: ...ii ...
Страница 4: ...iv ...
Страница 14: ...6 Chapter 1 Introduction ...
Страница 18: ...10 Chapter 2 Specifications ...
Страница 36: ...28 Chapter 3 Safety ...
Страница 46: ...38 Chapter 4 System Overview ...
Страница 54: ...46 Chapter 5 Troubleshooting ...
Страница 79: ...Chapter 6 Replacement Procedures 71 ...
Страница 80: ...72 Chapter 6 Replacement Procedures ...
Страница 82: ...74 Chapter 7 Maintenance ...
Страница 90: ...82 Chapter 8 Performance Testing ...
Страница 102: ...94 Appendix A Replacement Parts ...
Страница 109: ...101 ...
Страница 110: ...102 P25393 01 P25393 01 ...