![FujiFilm Sonosite PX System Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/fujifilm/sonosite-px-system/sonosite-px-system_service-manual_2342370047.webp)
Chapter 5:
Troubleshooting
39
Chapter 5: Troubleshooting
This chapter contains information to help you correct problems with system operation.
Note: If the system requires repair, certain steps must be taken to remove patient data from the system
prior to return to Fujifilm Sonosite. To accomplish this, the patient data needs to be wiped from the system
whenever possible (certain conditions may prevent this, such as a system that fails to power on). If unable
to power on, the Patient Data Drive can be removed as shown in
Figure 6.2 on page 48
.
There are two methods available to wipe patient data
. Full Factory Reset or ePHI Reset. The ePHI
Reset deletes all patient data and re-encrypts the drive. The Full Factory Reset performs a full reset of all
user, patient, preference data, re-encrypts the drives and starts the Out of Box Experience wizard.
Both
options are available to Administrators and users.
A third method is to install a properly configured
USB drive into the system.
Warning:
This will erase all patient data saved on the system and re-encrypt the patient drive. Patient
images should be exported or archived before proceeding.
Administrator
1
Using your administrative login information, log into the administrative settings page.
2
Tap
Admin Settings
.
3
Tap
Delete all patient data
for the ePHI reset or the
Factory Reset
button for Full Factory Reset
.
4
Make sure the system is connected to AC power, and tap
Yes
to continue. The process will take
approximately 35 minutes.
Note: If you are not connected to AC power and the batteries die, you will need to restart the process.
5
When the wipe is complete, tap
OK
to restart the system.
User
1
Simultaneously press the
AUTO
,
DEPTH DOWN
, AND
M
buttons for the ePHI Reset or
AUTO, DEPTH
DOWN,
AND
2D
buttons for the Full Factory Reset.
2
Make sure the system is connected to AC power, and tap
Yes
to continue. The process will take
approximately 35 minutes.
Note: If you are not connected to AC power and the batteries die, you will need to restart the process.
3
When the wipe is complete, tap
OK
to restart the system.
USB
If the user inserts a properly configured flash drive and then powers on the device, the system will start the
Full Factory Reset procedure. This mechanism is provided in order to allow the user to attempt recovering
the system in the event that the application has failed and is no longer able to start the reset via the
combination of keys.
Flash drive procedure
1
create a text file named “factoryreset.info” (do not include quotations) on the root of the drive.
2
Insert the flash drive at boot or after the application is running: in either case, the reset will start.
Содержание Sonosite PX System
Страница 1: ...i Sonosite PX Ultrasound System Service Manual ...
Страница 2: ...ii ...
Страница 4: ...iv ...
Страница 14: ...6 Chapter 1 Introduction ...
Страница 18: ...10 Chapter 2 Specifications ...
Страница 36: ...28 Chapter 3 Safety ...
Страница 46: ...38 Chapter 4 System Overview ...
Страница 54: ...46 Chapter 5 Troubleshooting ...
Страница 79: ...Chapter 6 Replacement Procedures 71 ...
Страница 80: ...72 Chapter 6 Replacement Procedures ...
Страница 82: ...74 Chapter 7 Maintenance ...
Страница 90: ...82 Chapter 8 Performance Testing ...
Страница 102: ...94 Appendix A Replacement Parts ...
Страница 109: ...101 ...
Страница 110: ...102 P25393 01 P25393 01 ...