Chapter 6:
Replacement Procedures
63
Required Part
• P27411-XX Service Assembly, ComE PCBA
Remove ComE PCBA
1
Use a T8 bit to remove the 4 screws securing the ComE PCBA to the Main PCBA
2
Gently lift the ComE PCBA using a slight “rocking motion” to release the connections
Figure 6.22 ComE PCBA removal
Install ComE PCBA
1
Place ComE PCBA onto Main PCBA. Make sure connectors are aligned and press into place
per
Figure 6.23 on page 63
.
2
Use a T8 bit to install 4 screws and tighten to 5.5 inch pounds.
Figure 6.23 Install ComE PCBA
Remove 4 screws
Ensure connectors
align properly