2006. Mar
53
2. Warnings
Laser error occurred.
Use other lasers.
Failed to retrive picture data from
PC.
Check the PC setting.
PMT error occurred.
Use low sensitivity for the setting.
The file name is already used.
Close the SampleSetDoor or
FilterChangeDoor.
Meaning:
• A laser error has occurred.
• Use a different laser.
Countermeasure:
• An error has occurred with the
laser.
• If an error occurs even after
restarting the instrument, call a
serviceman.
• It is possible to scan using a
different laser.
Meaning
• There was a failure in reading the
image data from the computer.
• Check the computer environment.
Countermeasure:
• It is possible that the computer's
processing ability has degraded.
• Check the computer once, when
there are no devices connected to
it and no software started.
Meaning:
• An overexposure error has
occurred.
• Scan with low sensitivity settings.
Countermeasure:
• Overexposure has been detected
outside of the scanning area.
• If an error occurs even after
restarting the instrument, call a
serviceman.
Meaning:
Close the sample set door or the
filter change door.
Countermeasure:
Close the sample set door or the
filter change door.
Содержание FLA-7000
Страница 1: ...Version 2 0 2006 Mar Operation Manual...
Страница 2: ...2006 Mar...
Страница 5: ...Part 1 Preparations Before Use...
Страница 9: ...Part 2 Reading Imaging Plates...
Страница 17: ...Part 3 Reading Fluorescent Samples...
Страница 25: ...Part 4 Reading Digitized Samples...
Страница 32: ...28 2006 Mar...
Страница 33: ...Part 5 Lasers and Filters Other Settings...
Страница 44: ...40 2006 Mar...
Страница 45: ...Part 6 Installing and Uninstalling the Software...
Страница 55: ...Part 7 Troubleshooting...