7.3 Auto calibration (option)
Description
Calibration is performed at time intervals set in advance.
The solenoid valve is driven by contact signal to feed the standard gas for automatic calibration with
span gas and zero gas.
“CAL” is displayed on the left of the measurement screen during automatic calibration.
If the output signal hold is set, the output signal is held to the set value during calibration.
After the calibration, the hold is maintained until the time set in the measurement recovery time elaps-
es.
For automatic calibration, it is necessary to set “10.2.2 Date and time for starting automatic calibra-
tion (option)”, “10.2.3 Cycle time setting of automatic calibration (option)”, “10.2.7 Calibration
gas setting”, “10.2.8 Calibration waiting setting (option)”, and “10.5.12 Setting of measurement re-
covery time
Refer to Sections “4.2” and “4.3” for the wiring of solenoid valves.
Procedu
re
Operation
(example)
Setting the automatic calibration so that it is performed every four days from 13:00,
2015/02/25
Key operation
Description
Displayed message
(LCD)
(1)
Display the screen on the right in accordance with the key op-
eration summary and press the
key.
CALIBRATION MENU
SET AUTO CAL
(2)
Press
the
key.
The auto calibration valid/invalid setting screen appears.
SET AUTO CAL
AUTO CALIBRATION
(3)
Use the
key to select the auto calibration valid (YES).
Press the
key to set the value.
AUTO CALIBRATION
YES
NO
(4)
Press
the
key to set the value.
AUTO CALIBRATION
YES
Содержание ZKMA
Страница 1: ......
Страница 13: ...2 INZ TN2ZKMA E 2 OPERATING PARTS AND THEIR FUNCTIONS 2 1 Outline Drawing 1 ZKMA 2 ZKMB...
Страница 22: ...11 INZ TN2ZKMA E 4 2 2 Cable gland and input and output lines 1 ZKMA 2 ZKMB...
Страница 23: ...12 INZ TN2ZKMA E 4 2 3 Allocation of the terminal block...
Страница 24: ...13 INZ TN2ZKMA E 4 3 Wiring and piping diagram 4 3 1 1 Flow guide tube system...
Страница 25: ...14 INZ TN2ZKMA E 4 3 1 2 Flow guide tube system with valve...
Страница 26: ...15 INZ TN2ZKMA E 4 3 1 3 Ejector system...
Страница 27: ...16 INZ TN2ZKMA E 4 3 1 4 Ejector system with valve...
Страница 31: ...20 INZ TN2ZKMA E 5 2 Key operation flow diagram outline...
Страница 32: ...21 INZ TN2ZKMA E...
Страница 33: ...22 INZ TN2ZKMA E...
Страница 34: ...23 INZ TN2ZKMA E...
Страница 113: ...102 INZ TN2ZKMA E 11 SPECIFICATIONS 11 1 Specifications...
Страница 114: ...103 INZ TN2ZKMA E...
Страница 115: ...104 INZ TN2ZKMA E 11 2 Code symbols Converter...
Страница 116: ...105 INZ TN2ZKMA E Detector...
Страница 117: ...106 INZ TN2ZKMA E Ejector Exclusive cable...
Страница 118: ......