- 49 -
INZ-TN2ZKMA-E
9.5 Troubleshooting
Phenomena
Probable causes
Checking methods (normal value)
Remedy
No display
Converter fuse blown out
Check the fuse and supply voltage speci-
fication.
Replace fuse Check
Power supply voltage
Indication does
not change or
slow response
Filter and/or flow guide tube
clogged
Visual check of filter and flow guide
tube for contamination or clogging.
Check for loosen and gas leaks at piping
connections and mounting place of de-
tector.
Clean or replace filter
Tighten pipe connec-
tions
Sensor deterioration
Change over between zero and span gas
and check if 5 minutes or longer is need-
ed for 90% response.
Replace sensor
Decrease in flow velocity of
exhaust gas
Check response to process gas after
shutting down calibration gas.
Move the direction (mounting position)
of “arrow” of the flow guide slightly.
Increase process gas
Flow into the flow
guide tube.
Temperature
alarm contin-
ues for more
than 10 min.
after power
switched ON
Break of wiring
Wrong wiring
Source voltage is too low.
Cable check of wiring
Wiring check
Check of supply voltage specification
Replacement
Correct wiring
Check power supply
Break of thermocouples
Break check
Replace sensor
Blown heater fuse
Cable check of fuse
Replace fuse
Break in detector heater
Check heater resistance
50 to 55
Ω
for 115V, 200 to 250
Ω
for
220V (Excluding wiring resistance)
Replace sensor
Automatic
calibration is
not possible
Difference between calibration
gas concentration and its setting
Check the set value for calibration gas
concentration.
Set proper value (Re-
fer to “10.2.7”)
Wrong parameters setting
Check automatic calibration intervals.
Set proper parame-
ters
The calibration is prohibited in
the contact input of the external
terminal block.
Check if the calibration is not prohibited
in the contact input of the external ter-
minal block.
Set proper parame-
ters
Correct wiring
The heater is set to off at the
contact input of the external
terminal block.
Check if the heater is set to off at the
contact input of the external terminal
block.
Set proper parame-
ters
Correct wiring
Zero and/or
span alarm
Difference between calibration
gas concentration and its setting
or misconnection between zero
and span gas
Check the set value for calibration gas
concentration.
Set proper value
Check piping.
Correct wiring
Indication too
high or too low
Loose flange and its surround-
ings Deteriorated O-rings
Check for gas leaks in detector and
mounting part of flow guide tube flange.
Tighten mounting
screws
Replace sensor
Check for leaks from the outside.
Seal
Detector is faulty.
Check for gas leaks at calibration gas
inlet.
Check sensor voltage (mV) for higher or
lower than other detector when flowing
zero gas. (See “6.5 Oxygen detector
standard output voltage”)
Tighten connectors
Replace sensor
Abnormal detector temperature Refer to check items for detector tem-
perature alarm described above.
Replace sensor
Indication difference between
dry and wet base measurement
Oxygen concentration is higher in dry
base.
Normal
Unstable read-
ings
The power supply voltage is too
high or too low.
Check that the supplied voltage is as
specified.
Auto-tuning
Содержание ZKMA
Страница 1: ......
Страница 13: ...2 INZ TN2ZKMA E 2 OPERATING PARTS AND THEIR FUNCTIONS 2 1 Outline Drawing 1 ZKMA 2 ZKMB...
Страница 22: ...11 INZ TN2ZKMA E 4 2 2 Cable gland and input and output lines 1 ZKMA 2 ZKMB...
Страница 23: ...12 INZ TN2ZKMA E 4 2 3 Allocation of the terminal block...
Страница 24: ...13 INZ TN2ZKMA E 4 3 Wiring and piping diagram 4 3 1 1 Flow guide tube system...
Страница 25: ...14 INZ TN2ZKMA E 4 3 1 2 Flow guide tube system with valve...
Страница 26: ...15 INZ TN2ZKMA E 4 3 1 3 Ejector system...
Страница 27: ...16 INZ TN2ZKMA E 4 3 1 4 Ejector system with valve...
Страница 31: ...20 INZ TN2ZKMA E 5 2 Key operation flow diagram outline...
Страница 32: ...21 INZ TN2ZKMA E...
Страница 33: ...22 INZ TN2ZKMA E...
Страница 34: ...23 INZ TN2ZKMA E...
Страница 113: ...102 INZ TN2ZKMA E 11 SPECIFICATIONS 11 1 Specifications...
Страница 114: ...103 INZ TN2ZKMA E...
Страница 115: ...104 INZ TN2ZKMA E 11 2 Code symbols Converter...
Страница 116: ...105 INZ TN2ZKMA E Detector...
Страница 117: ...106 INZ TN2ZKMA E Ejector Exclusive cable...
Страница 118: ......