INF-TN2FSC-E
43
(2) Image figure of mounting dimension
FSSE
FSSC, FSSD, FSSH
Type
V method
V method
Z method
Z method
Mounting dimensions
Mounting
dimensions
Mounting
dimensions
Mounting dimensions
Mounting
method
Mounting
dimensions
Type
Mounting
method
Mounting
dimensions
FSSC, FSSD, FSSH
FSSE
(3) Detector selection standards
The Z method for large size sensor is recommended for outer diameter 300mm or more.
FSSE should be used as much as possible for pipes such as old pipes, cast iron pipes, and mortar
lining pipes, through which it is difficult for ultrasonic signals to pass.
Z
V
FSSE
FSSD
V
FSSH
Z
V
Z
V
FSSC
-40 to 120
-40 to 80
-40 to 100
-40 to 200
13 25
50
100
200
300
400
1000
3000
6000
250
Type
Fluid
temperature [°C]
Inner diameter of piping ø (mm)
Px
Px, P, M
Px, P, M
Px, P, M
P, M
Px, P, M
Px, P, M
300
13
Z
P, M
300
150
50
50
150
400
250
200
1200
Px
200
600
200
3000
6000
200
300
P, M
50
600
Px : PP, PVDF
P : Plastic (PVC, etc.)
M : Msetallic piping (steel pipe, copper pipe, aluminum, etc.)
Classification of
piping materials
600