136
INF-TN2FSC-E
10.7 Flow velocity distribution display function (optional)
It is possible to measure the flow velocity distribution in real time by the pulse doppler method
and to display the flow state in the piping.
Use this function for judgment if the flow rate measuring position is appropriate, for diagnosis
of flow, for research, testing and others.
This function is applicable to the following types.
Main unit type: FSC
1
-
(with flow velocity distribution display function, if the 6th digit is "1")
Sensor type: FSDP2 (Bore diameter;
φ
40 to 200mm Fluid temperature; –40 to +100ºC)
FSDP1 (Bore diameter;
φ
100 to 400mm Fluid temperature; –40 to +80ºC)
FSDP0 (Bore diameter;
φ
200 to 1000mm Fluid temperature; –40 to +80ºC)
10.7.1 Installing Detector
(1) Processing of detector mounting surface
Remove rust, pitch, surface irregularity and others from the pipe surface, to which a sensor is to
be mounted, by the frame length of the sensor to be mounted, using thinner, sandpaper and/or
other appropriate means.
Note) If the piping outer circumference is wound with jute, remove the jute from the entire outer
circumference in a length that is frame length (L) + 200 mm.
Wound with jute
Piping
L
+
200mm
(2) Installation of detector
(1) Wrap the belt around the pipe.
FSDP2, FSDP1
Adjusts the length of the wire rope according to the
piping size, fixes the wire on the pipe.
FSDP0
(2) Fully screw up to the right side.
Screw