POWER
The Bubbler requires a ded12VDC power source to operate. Power supplied
from dataloggers is insufficient to run the compressor.
IMPORTANT!
If not using an FTS supplied power cable (Part #18945)
between the battery and the Bubbler, an equivalent power cable
should be used. The power cable should be 14 AWG to a maximum of
8 ft with an integrated 20A slow blow fuse.
To connect the power cable:
1) Remove the power terminal block from the port
2) Secure the power wires to the appropriate pins (positive
and negative)
3) Connect the power cable to the power source (positive
to positive, negative to negative).
4) To apply power to the Bubbler, re-insert and secure the
power terminal block.
CONFIGURING THE BUBBLER
DEFAULT SETTINGS
The Bubbler is shipped with the following default settings:
SDI-12 Address
0
Sample period
8 secs
Bubbler rate/minute 60
Temperature
°F
Purge Pressure
90 psi
4mA (min)
0 feet
Stage Units
Feet
20mA (max)
69.2 feet