CONFIGURING USING OTHER DATA LOGGERS 

Use SDI-12 commands to change the Bubbler’s default settings.  Confirm 
communication with the sensor, by entering the Send Identification command. 

To use SDI-12 commands with an Axiom Datalogger enter Transparent Mode.  
Select SDI-12 >          which appears at the bottom of the screen.

 

 

SDI-12 COMMANDS 

The FTS Bubbler complies with SDI-12 Standard version 1.3 communication 
protocol and standard commands and responses.  Replace the 

in all SDI-12 

commands with the Bubbler’s address.  Some common SDI-12 commands are 
listed here. 

?!

 

Address 
query

 

Response will be 
the SDI-12 address

 

 

a

I

!

 

Send  
Identification

 

Responds with 
identifying 
information on 
sensor at the input 
address

 

a

Ab!

 

Change 
address

 

b

=the new address

 

 

 

MEASUREMENT COMMANDS 

See the X-Command table for stage units, temperature and purge state codes 

aM! 
aC! 
aMC! 
aCC! 

Measures:  

 

Stage 

 

Units of stage 

 

Status  

 

aM2! 
aC2! 
aMC2! 
aCC2! 

Measures:  

 

Purge state* 

 

Pump on time 

 

Pump on-off count 

 

Pump fault count 

 

Uncompensated stage 

 

Status 

aM1! 
aC1! 
aMC1! 
aCC1! 

Measures: 

 

Stage 

 

Units of stage* 

 

Temperature 

 

Units of Temperature* 

 

Offset (Units same as Stage) 

 

Low accuracy Line Pressure (psi) 

 

Tank Pressure (psi) 

 

Battery Voltage (volts) 

 

aM7! 
aC7! 
aMC7! 
aCC7! 

Measures: 

 

Purge state 

 

Pump on time 

 

Pump on-off count 

 

Pump fault count 

 

Uncompensated stage 

 

Status 

 

 

10 

Содержание Bubbler

Страница 1: ...EXTREME ENVIRONMENTS EXTREMELY RELIABLE 700 Bubbler QS Rev 1 11 Dec 2019 Part 20792 BUBBLER Installation and Quick Start Guide...

Страница 2: ...ions Connecting the Desiccator Orifice Line Power 2 3 3 4 5 6 CONFIGURING THE BUBBLER Default Settings Configuring Using the Bubbler s Front Panel Bubbler s Menu Tree Configuring with Axiom Datalogger...

Страница 3: ...olt to a vertical surface using the corner mounting plates 3 Connectors and outlets should point down DESICCATOR USE 1 A desiccator MUST be used on the Bubbler s air inlet line 2 The entire air intake...

Страница 4: ...orifice line to aid in securing the line in a fixed position Install the line outlet perpendicular to the flow in flowing water Secure the terminal end of the orifice line in a fixed position and with...

Страница 5: ...Connections Pin Description Pin Description 1 12 VDC 3 Ground 2 Serial data 4 4 20mA positive 3 Ground 5 4 20mA negative 6 Chassis ground 6 Chassis ground 1 2 3 4 5 6 PINS Barbed hose fitting for des...

Страница 6: ...PPLIED DESICCATOR 1 Mount the desiccant bracket 2 Remove the shipping plugs from the inlet and outlet sockets 3 Wrap Teflon tape around the threads of the barbed desiccant adaptor fitting Part 19469 S...

Страница 7: ...lace the nut and ferrules on the line as shown and fully insert the line into the Bubbler s outlet fitting 2 Screw the nut on until finger tight 3 Mark the nut at the 6 o clock position 4 Use a wrench...

Страница 8: ...use To connect the power cable 1 Remove the power terminal block from the port 2 Secure the power wires to the appropriate pins positive and negative 3 Connect the power cable to the power source posi...

Страница 9: ...MENUS PRESSURE PURGE BLEED STAGE UNITS 4mA 20mA Output mA SLOPE SITE FAULT STAGE OFFSET TANK PSI PURGE SET STAGE SETUP DIAGNOSTICS SITE ADDRESS RATE TEMP F C AVG DUR SDCARD BATTERY TEMP VERSION UPDATE...

Страница 10: ...o use a pre configured Datalogger or load a Configuration file from a USB stick 1 Insert the USB stick into one of the USB HOST ports on the Datalogger and select Station About Load Configuration 2 Co...

Страница 11: ...Home Sensors Bubbler Icon Once a valid reading has been returned select the Set Stage button and enter the stage reading 4 Set up Telemetry Logging Data Set up cog Add and other desired Features NOTE...

Страница 12: ...d here Address query Response will be the SDI 12 address aI Send Identification Responds with identifying information on sensor at the input address aAb Change address b the new address MEASUREMENT CO...

Страница 13: ...efault 10 8 aXBVLP p Get set battery voltage low P p is the minimum voltage that will trigger a battery fault Default 8 0 aXDOM Get date of manufacture Format is YYYYMMDD aXDTYYYYMMDD HHMMSS Get Set d...

Страница 14: ...et is set to zero does not zero the stage reading only the offset aXSLP Get set slope Range 0 100 999 999 Used when the site requires non standard stage unit See X command aXSUP for the built in list...

Страница 15: ...ample takes approximately 2 seconds aXSUP Get Set stage units 0 ft feet slope 2 307 1 in inches slope 27 680 2 m meters slope 07 03 3 cm centimeters slope 70 3250 4 mm millimeters slope 703 250 5 cu C...

Страница 16: ...tubing to the Bubbler 19443 1 5m SDI communication cable with flying leads both ends 19442 1 5m SDI communication cable MC one end and flying leads other end 18945 8ft cable with 20A fuse flying lead...

Отзывы: