CONFIGURING USING OTHER DATA LOGGERS
Use SDI-12 commands to change the Bubbler’s default settings. Confirm
communication with the sensor, by entering the Send Identification command.
To use SDI-12 commands with an Axiom Datalogger enter Transparent Mode.
Select SDI-12 > which appears at the bottom of the screen.
SDI-12 COMMANDS
The FTS Bubbler complies with SDI-12 Standard version 1.3 communication
protocol and standard commands and responses. Replace the
a
in all SDI-12
commands with the Bubbler’s address. Some common SDI-12 commands are
listed here.
?!
Address
query
Response will be
the SDI-12 address
a
I
!
Send
Identification
Responds with
identifying
information on
sensor at the input
address
a
Ab!
Change
address
b
=the new address
MEASUREMENT COMMANDS
See the X-Command table for stage units, temperature and purge state codes
aM!
aC!
aMC!
aCC!
Measures:
•
Stage
•
Units of stage
•
Status
aM2!
aC2!
aMC2!
aCC2!
Measures:
•
Purge state*
•
Pump on time
•
Pump on-off count
•
Pump fault count
•
Uncompensated stage
•
Status
aM1!
aC1!
aMC1!
aCC1!
Measures:
•
Stage
•
Units of stage*
•
Temperature
•
Units of Temperature*
•
Offset (Units same as Stage)
•
Low accuracy Line Pressure (psi)
•
Tank Pressure (psi)
•
Battery Voltage (volts)
aM7!
aC7!
aMC7!
aCC7!
Measures:
•
Purge state
•
Pump on time
•
Pump on-off count
•
Pump fault count
•
Uncompensated stage
•
Status
10