background image

7)  Push the desiccant tubing firmly onto the dryer’s barbed desiccant adaptor fitting, 

ensuring a tight seal. 

8)  Attach the desiccant hose to the barbed hose fitting on the Bubbler’s air intake 

                                               

 
 
 
 
 
 

ORIFICE LINE 

I

nstall the orifice line following the IMPORTANT GUIDELINES section of this document.  

Connect the orifice line to the Bubbler’s outlet using the enclosed Swagelok fitting. 

 

USING THE SWAGELOK FITTING 

 

1)

 

Place the nut and ferrules on the line as shown and 
fully insert the line into the Bubbler’s outlet fitting.

 

 
 
 
 
 
 
 
 

2)

 

Screw the nut on until finger tight. 

3)  Mark the nut at the 6 o’clock position. 
 
4)  Use a wrench to tighten the nut one and a quarter 

turns (pass the 6 o’clock position once and stop at 
the  
9 o’clock position) 

 
DO NO

T

 

twist the Bubbler outlet.  Only turn the nut 

 

 

 

 

Shipping  
Plug 

 

IMPORTANT:

  To maintain a tight seal, FTS 

recommends using the provided Swagelok 
fitting with 

brass 

ferrules.  If using plastic 

ferrules, do not re-use them.

 

Desiccant 
Adaptor 
Part# 19469

 

 

AIRFLOW 

Low 
resistance 
inlet vent 
Part# 20625

 

 

Содержание Bubbler

Страница 1: ...EXTREME ENVIRONMENTS EXTREMELY RELIABLE 700 Bubbler QS Rev 1 11 Dec 2019 Part 20792 BUBBLER Installation and Quick Start Guide...

Страница 2: ...ions Connecting the Desiccator Orifice Line Power 2 3 3 4 5 6 CONFIGURING THE BUBBLER Default Settings Configuring Using the Bubbler s Front Panel Bubbler s Menu Tree Configuring with Axiom Datalogger...

Страница 3: ...olt to a vertical surface using the corner mounting plates 3 Connectors and outlets should point down DESICCATOR USE 1 A desiccator MUST be used on the Bubbler s air inlet line 2 The entire air intake...

Страница 4: ...orifice line to aid in securing the line in a fixed position Install the line outlet perpendicular to the flow in flowing water Secure the terminal end of the orifice line in a fixed position and with...

Страница 5: ...Connections Pin Description Pin Description 1 12 VDC 3 Ground 2 Serial data 4 4 20mA positive 3 Ground 5 4 20mA negative 6 Chassis ground 6 Chassis ground 1 2 3 4 5 6 PINS Barbed hose fitting for des...

Страница 6: ...PPLIED DESICCATOR 1 Mount the desiccant bracket 2 Remove the shipping plugs from the inlet and outlet sockets 3 Wrap Teflon tape around the threads of the barbed desiccant adaptor fitting Part 19469 S...

Страница 7: ...lace the nut and ferrules on the line as shown and fully insert the line into the Bubbler s outlet fitting 2 Screw the nut on until finger tight 3 Mark the nut at the 6 o clock position 4 Use a wrench...

Страница 8: ...use To connect the power cable 1 Remove the power terminal block from the port 2 Secure the power wires to the appropriate pins positive and negative 3 Connect the power cable to the power source posi...

Страница 9: ...MENUS PRESSURE PURGE BLEED STAGE UNITS 4mA 20mA Output mA SLOPE SITE FAULT STAGE OFFSET TANK PSI PURGE SET STAGE SETUP DIAGNOSTICS SITE ADDRESS RATE TEMP F C AVG DUR SDCARD BATTERY TEMP VERSION UPDATE...

Страница 10: ...o use a pre configured Datalogger or load a Configuration file from a USB stick 1 Insert the USB stick into one of the USB HOST ports on the Datalogger and select Station About Load Configuration 2 Co...

Страница 11: ...Home Sensors Bubbler Icon Once a valid reading has been returned select the Set Stage button and enter the stage reading 4 Set up Telemetry Logging Data Set up cog Add and other desired Features NOTE...

Страница 12: ...d here Address query Response will be the SDI 12 address aI Send Identification Responds with identifying information on sensor at the input address aAb Change address b the new address MEASUREMENT CO...

Страница 13: ...efault 10 8 aXBVLP p Get set battery voltage low P p is the minimum voltage that will trigger a battery fault Default 8 0 aXDOM Get date of manufacture Format is YYYYMMDD aXDTYYYYMMDD HHMMSS Get Set d...

Страница 14: ...et is set to zero does not zero the stage reading only the offset aXSLP Get set slope Range 0 100 999 999 Used when the site requires non standard stage unit See X command aXSUP for the built in list...

Страница 15: ...ample takes approximately 2 seconds aXSUP Get Set stage units 0 ft feet slope 2 307 1 in inches slope 27 680 2 m meters slope 07 03 3 cm centimeters slope 70 3250 4 mm millimeters slope 703 250 5 cu C...

Страница 16: ...tubing to the Bubbler 19443 1 5m SDI communication cable with flying leads both ends 19442 1 5m SDI communication cable MC one end and flying leads other end 18945 8ft cable with 20A fuse flying lead...

Отзывы: