700-SDI-RADAR-LX-80-Man Rev 1 08 Oct 2021
23
/49
X COMMAND DEFINITION
RESPONSE AND DETAILS
aXGDZ0!
Get minimum
active zone value
Response: a+f
In which “f” =the active zone minimum value.
Sensor will not report a measurement lower than this value.
f!
Set minimum
active zone value
Replace “f” with desired minimum active zone value.
Sensor will not report a measurement lower than this value.
aXGDZ1!
Get maximum
active zone value
Response: a+f
In which “f” = the active zone maximum value.
Sensor will not report a measurement higher than this value.
f!
Set maximum
active zone value
Replace “f” with desired active zone maximum value.
Sensor will not report a measurement higher than this value.
aXGSHR!
Get sensor height
value
Response: a+f
In which f = sensor height above riverbed
f!
Set sensor height
value
Replace “f” with sensor height above riverbed.
Response: a+f
f!
Set current staff
gauge reading
Response: a+f
In which “f” is the current staff gauge reading,
Device will calculate sensor height above riverbed as:
staff gauge r distance from sensor to water
aXGLAV!
Get average time
in second for level
Response: a+f
In which “f” = averaging time in seconds
f!
Set average time
in second for level
Replace “f” with desired averaging time in seconds.
Response: a+f
xGLFT!
Get filter type for
data filtering
Response: a+t
In which “t”= filter type:
0 – no filter
1 – IIR filter
2 – Moving average
3 – Median
4 – Standard deviation
xGLFT+t!
Set filter type for
data filtering
Replace “t” with desired filter type code (see codes above in
“Get filter type for data filtering”
Response: a+t
xGLIR!
Get IIR filter
constant
Response: a+c
In which “c” = IIR filter constant used in the IIR filter
xGLIR+c!
Set IIR filter
constant
Replace “c” with desired IIR filter constant used n the IIR filter.
Response: a+c
Recommended range is from +0.0 to +1.0
Содержание 21216
Страница 5: ...iii 10 1 SPECIFICATIONS 47 10 2 MECHANICAL ASSEMBLY 48...
Страница 39: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 34 49...
Страница 41: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 36 49...
Страница 42: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 37 49...
Страница 46: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 41 49 TROUBLESHOOTING...
Страница 47: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 42 49 continued...
Страница 48: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 43 49 continued...
Страница 49: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 44 49...
Страница 50: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 45 49...
Страница 51: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 46 49...
Страница 53: ...700 SDI RADAR LX 80 Man Rev 1 08 Oct 2021 48 49 10 2 MECHANICAL ASSEMBLY All measurements in mm...