168230.01EN_Version03

 FROMM

 

 

It is forbidden to duplicate this manual or parts of it in any way or any form without written permission of the author © 2006

 

 

 

SET-UP 

 

  To reach the set-up push and hold the DM button for 5 seconds. The display counts down  

 

from 5 till 0. 

  During count down push the TARE and PRINT button simultaneously. 

 

The display shows the software version followed by 

“SEt”

, the first item of the main menu. 

 
The main menu consists of: 
 

Set 

:  the parameters for scaling the indicator 

 

CALibr  :  the parameters for the weight calibration 

 

USEr 

:  the parameters for the display - and button settings 

 

oPtion 

:  the parameters for the RS232 output 

 

rtC 

:  the parameters for the optional real time clock 

 
REMARKS 

  With the arrow buttons it is possible to select the menu items, the parameters in the menu items 

and the values of the parameters. 

   To enter one of the above mentioned settings, the TARE (Enter) button must be used. 

  When a value must be giving in the least significant digit on the right is blinking. The blinking digit 

can be changed with the arrow buttons. With the DM/SD button the foregoing digit is going to 
blink. 

  When a setting is entered which is not wanted, the ON/OFF (ESC) button must be used. One step 

back by pushing once, two steps back by pushing twice and back to weighing by pushing three 
times. 

  When the jumper J3 is not placed only the rtC parameters can be reached. 

  The STI-7 indicator has the possibility to have a multirange readability. The steps are 1, 2 and 5 

and the point where the display switches, can be programmed under the item CAL. 

 

SCALING (SEt) 

 
Use the arrow button to select this menu item, push the TARE button to enter it. The parameters are: 
2.1a : Cnt-bY (= 

count 

by) 

readability per (1; 2; 5; 10; 20; 50) 

2.1b  :  dEC Pt  (= decimal point) (0 – 0.0000) 
2.1c  :  FS 

(= full scale) maximum weighing capacity (100 – 65535) 

2.1d  :  AdC 

(= A/D converter) internal value only to read, (maximum value 65535) 

2.1e  :  ZEro 

(= limit of the ZERO button)  (100 %; 4 % = -2 till +2) 

2.1f  :  Azt 

(= auto zero track)  (oFF; 0.6; 1.0; 3.0 display divisions) 

2.1g  :  StAb-L  (= stability level) limit in ADC parts for the motion indication (0 - 99) 

 

WEIGHT CALIBRATION (CALibr) 

 
Use the arrow button to select this menu item. Push the TARE button to enter it. This item has the 
same protection as the main menu. During count down push the TARE and PRINT button 
simultaneously. The display shows 

“CAL 2P” 

or 

“CALLIN”

 

ONLY ENTER THE ITEM CAL 2P!

 

The parameters are: 
2.2a  :  SCALEO  :  display reading at first calibration point (zero) 
2.2b  :  AdC 0 

:  ADC value of first calibration weight (point) 

2.2c  :  SCALE1  :  display reading at second calibration point (weight) 
2.2d  :  AdC 1 

:  ADC value of second calibration weight (point) 

2.1e  :  CbY-L1     :    countby level 1, the point where the readability switches to a higher step 
                                   

(multirange). When 0 is entered the multirange function is not used 

2.1f  :  CbY-L2     :    countby level 2, the point where the readability switches to second higher step 
                                      remark: it is not possible to switch from 5 to 10   
 
See chapter “Example of weight calibration”. 
REMARK: When count by is 1,2 or 5, the calibration weight must be at least 100 digits. 
                   When count by is 10, 20 or 50, the calibration weight must be at least 1000 digits. 

 

SET-UP 

Содержание FS 2 Series

Страница 1: ...is manual or parts of it in any way or any form without written permission of the author 2006 1 FROMM SEMI AUTOMATIC Weighting system Serie FS 2XX User manual Items No 16 8230 Version 01 soft S2009 0...

Страница 2: ...author 2006 2 INDEX Page 1 Index 2 3 2 General 4 3 Safety instructions and devices 5 6 4 Warnings 7 5 Main functions Operation 8 6 Set Up 9 10 7 Example weight calibration 11 8 Errors 12 9 Printer ou...

Страница 3: ...e Vous vous prot gerez ainsi et viterez des d t riorations sur votre appareil GB Important Read the operating instructions carefully before installation and before using this machine for the first tim...

Страница 4: ...Instructions and Warnings which point out the intended use and unsafe situations that could not be prevented in the design and manufacture of the weighting unit The duration of guarantee is provided t...

Страница 5: ...halt the machine immediately Only trained electricians may perform electrical work on the machine When installing the machine as a precaution always check that the controls and safety systems are corr...

Страница 6: ...extinguishers filled only with CO2 be used so as not to damage both the equipment on the machine and the wiring system In the event of flooding it is compulsory that all the power supplies be disconn...

Страница 7: ...operation or principles to the machine are fully responsible for those changes and or adaptations All relevant safety measures should be taken when carrying out any activity on the machine especially...

Страница 8: ...ners of the housing During calibration the functions are ON OFF ESC to switch the indicator on and off back to the previous step s ZERO arrow up to increase a value and to go upwards through the menu...

Страница 9: ...are 1 2 and 5 and the point where the display switches can be programmed under the item CAL SCALING SEt Use the arrow button to select this menu item push the TARE button to enter it The parameters a...

Страница 10: ...Option Use the arrow button to select this menu item push the TARE button to enter it The parameters are 2 4a Adr printer address 1 99 2 4b bAUd baudrate RS232 1200 2400 4800 9600 19200 2 4c LinEFd l...

Страница 11: ...splay shows AdC 0 ADC value of first calibration weight Press the ENTER button the display shows the last value When the scale is empty and stable press the F1 button the display shows the ADC value o...

Страница 12: ...7V LobAt 2 battery voltage transmitter is to low under 7V LobAt 3 battery voltage option board is to low under 12V ERR 01 the scale is too much negative under load ERR 02 the scale is far too much pos...

Страница 13: ...can be given in a number which is defined by the numbers in this list One or more lines can be selected by adding the value of the wanted line s 1 Id iiiii code number 2 Seq sss sequence number 4 Gro...

Страница 14: ...20 0D 0A Example 2 net and instable ASCII 65 example 2 Kg exam N CR LF Hex 20 20 20 2D 36 35 20 20 6B 67 20 4E 3F 0D 0A Example 3 gross 0 and stable ASCII 0 0 example 3 Kg exam CR LF Hex 23 20 20 20 3...

Страница 15: ...168230 01EN_Version03 FROMM It is forbidden to duplicate this manual or parts of it in any way or any form without written permission of the author 2006 15 JUMPER SETTINGS JUMPER SETTING...

Отзывы: