COMPUTER OUTPUT PROTOCOL

  

 
RS232 output   : 1 startbit, 8 databits, 1 stopbit. 
Baudrate           : selectable 1200 - 19200  in set-up item 2.4b (bAUd) 
DTR time-out    : 2 seconds see error 07 
 
This protocol only is valid if the automatic print function in 2.4d (AutoPr) is switched on and is not working 
after pressing the PRINT button. 
 
 

Field: 

Weight Space  Unit 

Space  Net 

Stable  CR 

LF 

Length: 

 

or 

1 3 1 1 1 1 1 

 
Field: 

Weight (gross or net) 

Space 

Unit 

Space 

Net 

Stable 

CR  LF 

Length: 

6 (7 with dec. Point) 

Example 1, gross and stable 

  

 

 

 

 

 

 

 

ASCII: 

 1244.5 (example 1) 

  

 Kg (exam)  

 

 

CR  LF 

Hex: 20,31,32,34,34,2E,35  20 20,6B,67 

20 20 20 0D 

0A 

Example 2, net and instable 

  

 

 

 

 

 

 

 

ASCII: 

    -65 (example 2) 

  

 Kg (exam)  

CR  LF 

Hex: 20,20,20,2D,36,35 

20 

20,6B,67 20 

4E 

3F 

0D 0A 

Example 3, gross 0 and stable 

  

 

 

 

 

 

 

 

ASCII: 

#   0.0 (example 3) 

  

 Kg (exam)  

  

  

CR  LF 

Hex: 23,20,20,20,30,2E,30  20 20,6B,67 

20 20 20 0D 

0A 

 
Weight  = 

Display weight lined up right, length depending on decimal point. 

Space 

=   ASCII space (20hex). 

Unit 

In the set-up given in unit (Unit). 

Net 

For the gross reading a space, for the net reading a letter "N". 

Stable 

If stable a space, if in motion a "?". 

CR 

=   Carriage return (0Dhex) 

LF 

Line Feed (0Ahex) 

 
The automatic printing is active only in the “gross” or  “net” mode.  
 
When by 2.4d (AutoPr) Option 4 is selected, print after DTR input is high, the next connection is valid: 
 
PTXD – data output indicator 

 computer (nr. 4 on RS232 connector – yellow) 

PDTR – data input indicator 

 computer (nr. 2 on RS232 connector – red) 

GND – ground, data input / output 

 computer (nr. 6 on RS232 connector – blue) 

 
Connection terminal mounted in housing base. 

 

COMPUTER OP PROTOCOL 

Содержание FS 2 Series

Страница 1: ...is manual or parts of it in any way or any form without written permission of the author 2006 1 FROMM SEMI AUTOMATIC Weighting system Serie FS 2XX User manual Items No 16 8230 Version 01 soft S2009 0...

Страница 2: ...author 2006 2 INDEX Page 1 Index 2 3 2 General 4 3 Safety instructions and devices 5 6 4 Warnings 7 5 Main functions Operation 8 6 Set Up 9 10 7 Example weight calibration 11 8 Errors 12 9 Printer ou...

Страница 3: ...e Vous vous prot gerez ainsi et viterez des d t riorations sur votre appareil GB Important Read the operating instructions carefully before installation and before using this machine for the first tim...

Страница 4: ...Instructions and Warnings which point out the intended use and unsafe situations that could not be prevented in the design and manufacture of the weighting unit The duration of guarantee is provided t...

Страница 5: ...halt the machine immediately Only trained electricians may perform electrical work on the machine When installing the machine as a precaution always check that the controls and safety systems are corr...

Страница 6: ...extinguishers filled only with CO2 be used so as not to damage both the equipment on the machine and the wiring system In the event of flooding it is compulsory that all the power supplies be disconn...

Страница 7: ...operation or principles to the machine are fully responsible for those changes and or adaptations All relevant safety measures should be taken when carrying out any activity on the machine especially...

Страница 8: ...ners of the housing During calibration the functions are ON OFF ESC to switch the indicator on and off back to the previous step s ZERO arrow up to increase a value and to go upwards through the menu...

Страница 9: ...are 1 2 and 5 and the point where the display switches can be programmed under the item CAL SCALING SEt Use the arrow button to select this menu item push the TARE button to enter it The parameters a...

Страница 10: ...Option Use the arrow button to select this menu item push the TARE button to enter it The parameters are 2 4a Adr printer address 1 99 2 4b bAUd baudrate RS232 1200 2400 4800 9600 19200 2 4c LinEFd l...

Страница 11: ...splay shows AdC 0 ADC value of first calibration weight Press the ENTER button the display shows the last value When the scale is empty and stable press the F1 button the display shows the ADC value o...

Страница 12: ...7V LobAt 2 battery voltage transmitter is to low under 7V LobAt 3 battery voltage option board is to low under 12V ERR 01 the scale is too much negative under load ERR 02 the scale is far too much pos...

Страница 13: ...can be given in a number which is defined by the numbers in this list One or more lines can be selected by adding the value of the wanted line s 1 Id iiiii code number 2 Seq sss sequence number 4 Gro...

Страница 14: ...20 0D 0A Example 2 net and instable ASCII 65 example 2 Kg exam N CR LF Hex 20 20 20 2D 36 35 20 20 6B 67 20 4E 3F 0D 0A Example 3 gross 0 and stable ASCII 0 0 example 3 Kg exam CR LF Hex 23 20 20 20 3...

Страница 15: ...168230 01EN_Version03 FROMM It is forbidden to duplicate this manual or parts of it in any way or any form without written permission of the author 2006 15 JUMPER SETTINGS JUMPER SETTING...

Отзывы: