FROMM
168230.01EN_Version03
It is forbidden to duplicate this manual or parts of it in any way or any form without written permission of the author © 2006
8
MAIN FUNCTIONS
The calibration is done only by the pushbuttons at the front side of the indicator. The calibration is
protected by jumper J3 (write protect) at the top on the right side of the main PCB. To enable to
calibration place this jumper at the pins. See jumper settings.
THE PUSHBUTTONS
The indicator has 7 pushbuttons. The most important ones are red coloured and can be found at the
corners of the housing. During calibration the functions are:
ON/OFF (ESC)
: to switch the indicator on and off, back to the previous step(s).
ZERO (arrow up)
: to increase a value and to go upwards through the menu items.
PRINT (arrow down)
: to decrease a value and to go downwards through the menu items.
TARE (Enter)
: to enter a setting.
F1
: function button to find the lowest value.
DM/SD
: to enter the menu and to select the blinking digit.
F2
:
no
function.
OPERATION
•
Connect the supply to the indicator. When a 230 VAC-supply is used, connect the adaptor.
•
Press the ON/OFF button, the display lights up.
•
Push the ZERO button when the reading is not zero, the reading becomes
“0”
and the
“ZERO”
indication left in the display lights up. The scale is now ready to use.
MAIN FUNCTIONS