- 13 -
2.
Unhook the eccentric lever from its position
In the presence of
liquid residues
, rotate the machine using the handle. For the model RS120 it is necessary
to unlock the machine with the knob on the side of the machine. The machine can be locked with the same
knob in the unloading position.
In the presence of
solid residue
, remove the retaining ring and remove the Rec-Bag inside the boiler paying
attention not to break it. The small amount of solvent that might have formed on the bottom of the boiler
can be removed like described above.
5.9.
Important advice
Do not rotate or shake the unit once loaded or when operating.
Clean the oil expansion vessel only with a 'wet' rag to avoid generating sparks.
Opening the cover sooner than one hour after the distillation cycle has finished will cause the cover gasket to
swell.
Some solvents during the distillation phase create such a quantity of foam that a correct separation of the sol-
vent from the polluting product is not possible. In these cases the distillate will be dirty. This inconvenience
can be overcome by using the optional antifoam kit.