Operation
Printing the Analyzer Report Header
2
2-33
Fluke Biomedical
Control #:
Serial #:
Model #:
Mfr.
Location:
Technician :
Date:
Settin
g
Actual
---------
20.0
J
Qwave Sync Time
43.9 ms
Rwave Sync Time 73.1
ms
Ener
g
y
19.9
J
Over Volta
g
e
0
V
Over
Current
0
A
Ener
g
y:
19.9
J
PACER OUTPUT MEASUREMENTS
Pulse
Rate:
40
PPM
Pulse
Width:
19.7
ms
Peak Current:
1
8
.0 ma
Peak Volta
g
e: 0.9
V
Ave
Current
0.0
ma
Ave Volta
g
e:
0.0
V
PACER REFRACTORY MEASUREMENTS
Selected Pacin
g
Rate:
40 PPM
Pulsed Refractory Period: 329.2 ms
Sensed Refractory Period: 179.2 ms
Char
g
e
Time: 05
sec
Ener
g
y:
19.
8
J
Header
fcf050.eps
Figure 2-10. Manual Output with Header