![Fluke QED 6 Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/fluke/qed-6/qed-6_user-manual_2310046039.webp)
Operation
Evaluating Ability to Fire
2
2-13
OFF
DEFIB
Defibrillator/Pacer
PACER
QED-6
DEFIBRILLA
TOR A
NA
LYZER
STERNUM (-)
BAT
TE
RY / 9VD
C
RS-232 / 10101
ECG
V
C
LL
F
LA
L
RL
N
RA
R
DEFIB
APEX (+)
PA
CE
R
1000 j
100 j
SOFTKEYS
fcf010.eps
Figure 2-7. ECG Lead Configuration
3.
From the
Energy
menu, press the
WAV>
softkey until the
VFIB
option
appears; then press the
VFIB
softkey.
A ventricular fibrillation waveform is simulated by the Analyzer through
the ECG jacks and paddle plates. When a discharge is complete, the
Analyzer outputs a 90 BPM Normal Sinus Rhythm.
4.
Press the
WAV>
softkey until the
VTACH
option appears; then press the
VTACH
softkey.
A ventricular tachycardia is simulated by the Analyzer through the ECG
jacks and paddle plates. When the discharge is complete, the Analyzer
outputs a 90 BPM Normal Sinus Rhythm.
Содержание QED 6
Страница 6: ......
Страница 10: ...QED 6 Users Guide iv ...
Страница 12: ...QED 6 Users Guide vi ...
Страница 14: ...QED 6 Users Guide 1 2 ...
Страница 28: ...QED 6 Users Guide 2 2 ...
Страница 32: ...QED 6 Users Guide 2 6 fcf018 eps Figure 2 2 Main Menu 1 Functions ...
Страница 33: ...Operation Powering Up 2 2 7 fcf003 eps Figure 2 3 Main Menu 2 Functions ...
Страница 34: ...QED 6 Users Guide 2 8 fcf004 eps Figure 2 4 Autosequence Menu Structure Main Menu 2 ...
Страница 68: ...QED 6 Users Guide 3 2 ...