Operation
Pacemaker
(Non-Invasive)
Testing
2
2-23
QED-6
DEFIBRILLATOR ANALYZER
STERNUM (-)
BATTERY / 9VDC
RS-232 / 10101
ECG
V
C
LL
F
LA
L
RL
N
RA
R
DEFIB
APEX (+)
PACER
1000 j
100 j
SOFTKEYS
O
u
tp
u
t
Pacer
Adapters
Pacer
ECG Leads
RA
LA
LL
fcf015.eps
Figure 2-9. Connecting Pacemaker Output to Analyzer
Note
The Pacer can be in either demand or non-demand mode.
5.
Press the
MEAS
softkey to display the following:
fcf031.eps
Three hyphens (---) indicate that no pacer pulses were received.
6.
Set the pacer at various current and heart rate settings. The results are
displayed. Press the
softkey for a hard copy.