2-22
4.
Press the two defibrillator paddles onto the contact electrode plates on the
front of the Analyzer.
5.
Press the
START
softkey and initiate the defibrillator charge cycle.
6.
As soon as the defibrillator reaches full charge, discharge it, noting the
time (in seconds) on the display. The maximum for the Analyzer is 60
seconds. After 60 seconds, the Analyzer displays
OVER
.
Pacemaker (Non-Invasive) Testing
1.
From
Main Menu 1
, press
more
to access
Main Menu 2
. The following
displays:
fcf026.eps
2.
Set the range switch to
pacer
. Otherwise the unit displays the following
message:
fcf029.eps
3.
From
Main Menu 2
, press the
PACE
softkey to display the
Pacer
Tests
menu:
fcf030.eps
4.
Connect the output from the pacer to the Analyzer, as shown in Figure
2-9.