![Fluke QED 6 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/fluke/qed-6/qed-6_user-manual_2310046052.webp)
QED 6
Users Guide
2-26
Programming an Automatic Test Sequence
The Model H can store in memory up to 28 automatic sequences to fully test
defibrillator performance according to protocol. Standard defaults for
Programs 0-27 for the Model H are listed in Table 2-3.
Table 2-3. Autosequencing Defaults (Model H)
Performance Parameter
Default
Print Heading
Yes
Energy Measurements
10 J
100 J
200 J
300 J
360 J
Energy / -
5%
Charge Time
Yes
Sync Time
Yes
Peak No
Overshoot No
ECG performance
30 BPM
120 BPM
240 BPM
2 Hz Sq
.125 Hz Sq
10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin
100 Hz Sin
2 Hz Tri
60 BPM PLS
Fib
Pacer No
Содержание QED 6
Страница 6: ......
Страница 10: ...QED 6 Users Guide iv ...
Страница 12: ...QED 6 Users Guide vi ...
Страница 14: ...QED 6 Users Guide 1 2 ...
Страница 28: ...QED 6 Users Guide 2 2 ...
Страница 32: ...QED 6 Users Guide 2 6 fcf018 eps Figure 2 2 Main Menu 1 Functions ...
Страница 33: ...Operation Powering Up 2 2 7 fcf003 eps Figure 2 3 Main Menu 2 Functions ...
Страница 34: ...QED 6 Users Guide 2 8 fcf004 eps Figure 2 4 Autosequence Menu Structure Main Menu 2 ...
Страница 68: ...QED 6 Users Guide 3 2 ...