QED 6
Users Guide
2-24
Note
Pacer voltage and current are displayed as average voltage and
current. If a printer is connected to the Analyzer, the printout also
documents peak voltage and current. If computer control is being
used, no peak values are available. All voltage measurements are
referenced to the internal 50
Ω
load.
Pacemaker Refractory Period Testing
To test the pacer refractory period:
1.
Set the pacer in demand mode.
2.
From
Main Menu 1
, press
more
to access
Main Menu 2
. The following
displays:
fcf026.eps
3.
Set the range switch to
pacer
. Otherwise the unit beeps and displays the
following message:
fcf029.eps
4.
From
Main Menu 2
, press the
PACE
softkey to display the
Pacer
Tests
menu:
fcf032.eps
5.
Press
RFP
to select refractory period testing. Refer to Figure 2-9 for
setup. The following menu appears.