Remote Control - Description of Commands
Commands and Queries
4
4-41
Example(s)
VOLT3:RANG:AUTO ON
CURR2:RANG:AUTO? Response:
0
*RST state
ON
Invalidates
[SENSe:]CURRent[1..6]|VOLTage[1..6]:RANGe[:UPPer]
Invalidated by
[SENSe:]CURRent[1..6]|VOLTage[1..6]:RANGe[:UPPer]
[SENSe:][:POWer[1..6]]|CURRent[1..6]|VOLTage[1..6]:AC[|:DC]:APERture[:TIME]
<avgtime>
Nominal averaging interval
Actual averaging interval
Signal on synchronization source (INP[1..12] or EXTernal
Description
Sets the nominal averaging interval. Query returns the set nominal averaging interval. In
synchronous mode the actual averaging interval is changing “on-the-fly”. The nominal
averaging interval is extended to next full signal period.
To query the actual averaging period, :SENSe:DATA? "TIME[:INTerval]" command
must be used.
If the nominal averaging interval is changed by this command, then the synchronization
timeout (SYNC:TIMeout) is set to the nominal averaging interval or 0.3 seconds,
whichever is greater.
Note
The [:POWer[1..6]]|CURRent[1..6]|VOLTage[1..6]:AC[|:DC] nodes are
implemented for SCPI compatibility only. The instrument works with only
one averaging interval for all averaged measurements.
Parameter(s)
15ms ... 3600 s
Resolution is 1 ms.
The unit is seconds.
Example(s)
APER 0.2
APER? Response:
1.5
Содержание NORMA 4000 Operators
Страница 6: ...Remote Control Users Guide iv ...
Страница 8: ...Remote Control Users Guide vi ...
Страница 10: ...Remote Control Users Guide viii ...
Страница 12: ...Remote Control Users Guide 1 2 Output Unit 1 13 Command Sequence and Command Synchronization 1 14 ...
Страница 26: ...Remote Control Users Guide 2 2 ...
Страница 38: ...Remote Control Users Guide 3 2 ...
Страница 50: ...Remote Control Users Guide 4 2 ...
Страница 150: ...Remote Control Users Guide 4 102 ...
Страница 152: ...Remote Control Users Guide 5 2 ...
Страница 160: ...Remote Control Users Guide 6 2 ...
Страница 184: ...Remote Control Users Guide 6 26 ...