
Remote Control - Description of Commands
Commands and Queries
4
4-33
INPut[1..12]:COUPling AC | DC
Description
Sets the input coupling of the selected input channel.
Parameter(s)
AC
DC component is removed from the signal before any further processing. This coupling
compensates for any DC offsets on the signal.
DC
Signal is left untouched and all it’s components are passed for further processing. This
coupling should be used for true RMS calculations.
*RST state
DC
for all channels
Example(s)
INP1:COUP AC
INP2:COUP? Response:
DC
Invalidates
-
Invalidated by
-
INPut[1|2|3|4|5|6|7|8|9|10|11|12]:GAIN <gain>
Description
Sets the shunt factor for the voltage current inputs. This setting applies when EXTernal
input shunt is selected. The even (voltage) channel numbers are available only on
instruments equipped with PP59/PP69 power phase.
Parameter(s)
1.0e-7 to 1.0e+7
This unitless gain factor specifies V/A conversion ratio of the external shunt connected to
the specified channel. Negative values are not allowed.
Response
<gain>
*RST state
1.0 for all channels
Example(s)
INP1:GAIN 10.0
INP3:GAIN? Response:
251.65
Invalidates
-
Invalidated by
-
Содержание NORMA 4000 Operators
Страница 6: ...Remote Control Users Guide iv ...
Страница 8: ...Remote Control Users Guide vi ...
Страница 10: ...Remote Control Users Guide viii ...
Страница 12: ...Remote Control Users Guide 1 2 Output Unit 1 13 Command Sequence and Command Synchronization 1 14 ...
Страница 26: ...Remote Control Users Guide 2 2 ...
Страница 38: ...Remote Control Users Guide 3 2 ...
Страница 50: ...Remote Control Users Guide 4 2 ...
Страница 150: ...Remote Control Users Guide 4 102 ...
Страница 152: ...Remote Control Users Guide 5 2 ...
Страница 160: ...Remote Control Users Guide 6 2 ...
Страница 184: ...Remote Control Users Guide 6 26 ...