Remote Control
Users Guide
4-22
*RST state
-
Example(s)
CALC:INT:STOP
Invalidates
-
Invalidated by
-
CALCulate:INTegral:STOP:TIME <yyyy,MM,dd,hh,mm,ss>
Description
Specifies integration stop time. The integration stops when the instrument’s internal
date/time is equal to the time specified with this command.
Parameter(s)
yyyy Year
MM Month
Dd Day
Hh
Hours in 24 hour notation
Mm Minutes
ss
Seconds (integer value)
*RST state
2010,1,1,0,0,0
Example(s)
CALC:INT:STOP:TIME 2002,01,12,12,30,00
CALC:INT:STOP:TIME? Response:
2002,01,12,12,30,00
Invalidates
-
Invalidated by
-
CALCulate:INTegral:STOP:TINTerval <interval>
Description
Specifies integration interval in seconds. The integration after this interval elapses from
the start of the integration.
Parameter(s)
1.0e-3 to 9.99e+6
*RST state
6.01