Remote Control
Users Guide
4-62
SOURce:VOLTage[1..4][:LEVel][:IMMediate][:AMPLitude] <value>
Description
Selects the output voltage for FIXed mode.
Parameter(s)
<value>
Valid Range: -10.3 to 10.3 V
*RST state
0.0
Example(s)
SOUR:VOLT4 5.3
SOUR:VOLT1? Response:
-2.5
Invalidates
-
Invalidated by
-
SOURce:VOLTage[1..4]:FEED <function>
Description
Specifies the output reference function for VARiable mode.
Parameter(s)
<function>
Any valid averaged measurement function of the instrument.
*RST state
"VOLTage1"
Example(s)
SOUR:VOLT2:FEED "POW2:APP"
SOUR:VOLT4:FEED? Response: "CURR3:MEAN"
Invalidates
-
Invalidated by
-
Scaling
SOURce:VOLTage[1..4]:GAIN <value>
Description
Specifies the scaling for the output. The difference between the actual value of the
reference function and the ZERO value is multiplied by this factor to calculate the output
voltage.
Содержание NORMA 4000 Operators
Страница 6: ...Remote Control Users Guide iv ...
Страница 8: ...Remote Control Users Guide vi ...
Страница 10: ...Remote Control Users Guide viii ...
Страница 12: ...Remote Control Users Guide 1 2 Output Unit 1 13 Command Sequence and Command Synchronization 1 14 ...
Страница 26: ...Remote Control Users Guide 2 2 ...
Страница 38: ...Remote Control Users Guide 3 2 ...
Страница 50: ...Remote Control Users Guide 4 2 ...
Страница 150: ...Remote Control Users Guide 4 102 ...
Страница 152: ...Remote Control Users Guide 5 2 ...
Страница 160: ...Remote Control Users Guide 6 2 ...
Страница 184: ...Remote Control Users Guide 6 26 ...