Remote Control
Users Guide
1-4
4.
Go to File > Properties > Settings > ASCII Setup and check the following items:
•
Send line ends with line feeds
•
Echo typed characters locally
•
Append line feeds to incoming line ends
and press OK twice.
5.
In the main (white) window, type
*IDN?
and press
Enter
. (Do not make a
mistake while typing, since all characters you type are immediatelly sent to
instrument when you press a key on the keyboard. The backspace key will not
erase mistyped characters. If you make a mistake while typing, press Enter
several times. This will get things back in order).
6.
The instrument will return the identification string, for example:
Fluke,NORMA4000,KN34512BA,01.00
7.
In the main (white) window, type
DATA? “POW”
and press
Enter
. This will
instruct the instrument to return last valid power measurement.
8.
The instrument will return last valid power measurement, for example:
+1.02
Switchover to Remote Control
On power-up, the instrument is always in the manual control mode ("LOCAL" state) and
can be operated via the front panel.
The instrument is switched to remote control ("REMOTE" state) as follows:
IEC/IEEE-bus:
when it receives an addressed command from the controller with REN
line set.
Other interfaces: when it receives a valid command terminated by line feed <LF>
(=0Ah) from the controller in “SYSTem:KLOCk REM” state or by
this command explicitly.
During remote control, operation via the front panel is disabled. The instrument remains
in the remote state until it is reset to the manual state via the front panel or via the remote
control. Switching from manual to remote control and vice versa does not affect the
instrument settings.
Indications during Remote Control
The remote control state is indicated by a two-way radio icon in the leftmost cell of the
status line on the instrument’s screen. A key icon in the 3rd cell of the status line
indicates that the [LOCAL] key (F6/Esc) is disabled and switchover to manual control
can only be made via the remote control. If the key icon is not displayed, switchover to
manual control can be made with the [LOCAL] (F6/Esc) key.
Содержание NORMA 4000 Operators
Страница 6: ...Remote Control Users Guide iv ...
Страница 8: ...Remote Control Users Guide vi ...
Страница 10: ...Remote Control Users Guide viii ...
Страница 12: ...Remote Control Users Guide 1 2 Output Unit 1 13 Command Sequence and Command Synchronization 1 14 ...
Страница 26: ...Remote Control Users Guide 2 2 ...
Страница 38: ...Remote Control Users Guide 3 2 ...
Страница 50: ...Remote Control Users Guide 4 2 ...
Страница 150: ...Remote Control Users Guide 4 102 ...
Страница 152: ...Remote Control Users Guide 5 2 ...
Страница 160: ...Remote Control Users Guide 6 2 ...
Страница 184: ...Remote Control Users Guide 6 26 ...