Installation
Ethernet
33
5
5.9 Ethernet
The linescanner can communicate via Ethernet interface or RS485. During system configuration the user selects
either Ethernet or RS485. You cannot use Ethernet and RS485 communications for the data transfer at the same
time!
The Ethernet connection between linescanner and the PC has a maximum speed of 100 MBit/s and permits real-
time data transfer for all temperature pixels. For multi-scanner systems you can use a standard Ethernet Switch to
connect to the PC network adapter.
5.9.1 Connector
The LAN/Ethernet socket supports a M12 plug-in connector, 4 pin D-coded, and a screw retention feature.
Figure 5-2: M12 Connector Socket and Pin Assignment
For appropriate cables, see section 7.8
Ethernet Cable (A-CB-xx-M12-W04-xx)
5.9.2 Scanner Addressing
IP Address
The factory default IP address for the linescanner is 192.168.42.30.
The scanners address is not free of choice: It has to be unique in the network meaning that no other device in the
network including the PC network adapter may run at the same IP address.
Note
Ask your IT administrator for a free IP address to be used!
Advanced Ethernet Settings
Subnet Mask:
The Subnet Mask defines the interpretation of the IP address. The factory default setting is 255.255.255.0. The
Subnet Mask can be changed using the <NM> command of the scanner.
Port:
In the case that the default scanners port (2727) should conflict with something else (it could be blocked by the
firewall for example) it can be changed using the <PO> command of the scanner.
Gateway:
A gateway connects two subnets (which have a different subnet address). The IP address of this gateway is given
with the route command <RO>. The default is RO0.0.0.0 - this does mean no gateway is set.
Note
To establish the Ethernet communication, the subnet addresses for both the scanner and the PC need to
match! Appropriate changes can be applied either on the scanner side or on the PC network adapter!
M12-Pin
1
TX+
2
RX+
3
TX-
4
RX-
Содержание MP Series
Страница 16: ...MP Series Users Manual Rev 1 0 Apr 2021 16 1024 pixel 80 Hz scan rate Scan Angle FOV All models 90 ...
Страница 18: ...MP Series Users Manual Rev 1 0 Apr 2021 18 2 2 1 2 Far Focus Figure 2 2 Optical Diagrams Far Focus ...
Страница 37: ...Installation Ethernet 37 5 5 Close the dialog box by pressing on Save ...
Страница 44: ...MP Series Users Manual Rev 1 0 Apr 2021 44 7 2 Mounting Plate A MP MP Figure 7 1 Mounting Plate for Tripod ...